1. School of Biological Science and Medical Engineering, Southeast University, Nanjing 210096, China;2. Information Department, Jiangsu Province Hospital of Traditional Chinese Medicine, Nanjing 210029, China
The basic working principle of Cone Beam Computer Tomography (CBCT) is briefly described, the advantages and disadvantages of both continuous and pulsed X-ray imaging methods are analyzed, and the pulsed X-ray imaging method is adopted to implement the design of low-dose CBCT. On the basis of this, the internal and external trigger modes of PaxScan 2520 flat panel detector are realized. The experiment results show that the method of continuous X-ray imaging needs dozens of seconds for the process rotating a circle, in the meanwhile, the method of pulsed X-ray imaging only needs about four seconds for doing same job. In a word, pulsed X-ray imaging reduces the X-ray radiation time considerably and improves the safety performance of CBCT significantly. Besides, pulsed X-ray imaging with internal and external trigger modes has no difference in terms of image quality. The choice for the modes depends on whether or not needs active control for X-ray emission.
LUO Shouhua;CUI Yannan;BO Tengfei;LI Guang;CHEN Gong. , {{custom_author.name_en}}.
Control and Realization of Time Sequence for Low-dose Cone-Beam CT Imaging System[J]. Science & Technology Review, 2012, 30(5): 23-26 https://doi.org/10.3981/j.issn.1000-7857.2012.05.002