Articles

Possibility of Heart Rate Variability Analysis Using Ballistocardiogram

  • CAO Xinrong ,
  • WANG Kun ,
  • ZHANG Jing ,
  • TANG Jintian
Expand
  • 1. Institute of Medical Physics and Engineering, Tsinghua University, Beijing 100084, China;
    2. The Second Affiliated Hospital of Tsinghua University, Beijing 100049, China;
    3. School of Electrical Engineering, Hebei University of Technology, Tianjin 300401, China

Received date: 2013-07-23

  Revised date: 2013-09-12

  Online published: 2014-04-09

Abstract

Ballistocardiogram (BCG) is a non-contact cardiac activity monitoring method. Our preliminary research of heart rate variability (HRV) was carried out using BCG. BCG signal and electrocardiography (ECG) signal of 20 subjects were measured synchronously in static, valsalva maneuver, post valsalva maneuver and post exercise states. Cardiac periods of these two signals were calculated using RR intervals of ECG and JJ intervals of BCG respectively. HRV parameters in time domain analysis, in frequency domain analysis and nonlinear analysis were calculated. The correlation of HRVBCG parameters and HRVECG parameters were checked; the significant differences between HRVBCG and HRVECG parameters were tested using paired t-test. Results showed that the variance trends of cardiac periods calculated by ECG and BCG were the same; the correlations of HRVBCG and HRVECG parameters were all above 0.81; HRVBCG parameters showed no significant differences with HRVECG during static, post valsalva maneuver and post exercise state while nPNN50, LF/HF and SD1 showed significant differences during Valsalva maneuver state.

Cite this article

CAO Xinrong , WANG Kun , ZHANG Jing , TANG Jintian . Possibility of Heart Rate Variability Analysis Using Ballistocardiogram[J]. Science & Technology Review, 2014 , 32(4-5) : 86 -90 . DOI: 10.3981/j.issn.1000-7857.2014.h1.014

References

[1] Giovangrandi L, Inan O T, Wiard R M, et al. Ballistocardiography: A method worth revisiting[C]//Engineering in Medicine and Biology Society, EMBC. 2011 Annual International Conference of the IEEE, Boston, USA: IEEE, 2011(1): 4279-4282.
[2] Etemadi M, Inan O T, Giovangrandi L, et al. Rapid assessment of cardiac contractility on a home bathroom scale[J]. IEEE Transactions on Information Technology in Biomedicine, 2011, 15(6): 864-869.
[3] 陈灏珠. 心率变异性的测定及其临床意义[J]. 临床心电学杂志, 1994, 3 (4): 168-172. Chen Haozhu. Heart rate variability and the clinical significance[J]. Journal of Clinical Electroeardiology, 1994, 3(4): 168-172
[4] Acharya U R, Joseph K P, Kannathal N, et al. Heart rate variability: A review[J]. Medical and Biological Engineering and Computing, 2006, 44 (12): 1031-1051.
[5] 孙京霞, 白延强. 心率变异分析方法的研究进展[J]. 航天医学与医学工 程, 2001, 14(3): 230-234. Sun Jingxia, Bai Yanqiang. Research progress of methods of heart rate variabilty analysis[J]. Space Medicine & Medical Engineering, 2001, 14 (3): 230-234.
[6] 楚东雨, 宁新宝. 心率变异性的自回归模型功率谱分析[J]. 数据采集与 处理, 2000, 15(2): 259-262. Chu Dongyu, Ning Xinbao. PSD analysis of heart rate variability with ar model[J]. Journal of Data Acquisition & Processing, 2000, 15(2): 259- 262.
[7] Kamen P W, Tonkin A M. Application of the Poincaré plot to heart rate variability: A new measure of functional status in heart failure[J]. Australian and New Zealand Journal of Medicine, 1995, 25(1): 18-26.
[8] Brennan M, Palaniswami M, Kamen P. New insights into the relationship between Poincare plot geometry and linear measures of heart rate variability[C]//Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE, 2001(1): 526- 529.
[9] Cao X, Guo H, Tang J. Heart rate extraction of ballistocardiogram based on hilbert-huang transformation[C]//World Congress on Medical Physics and Biomedical Engineering, Beijing, China: Springer Berlin Heidelberg, 2013: 619-622.
[10] Gonzalez-Landaeta R, Casas O, Pallas-Areny R. Heart rate detection from an electronic weighing scale[J]. Physiological Measurement, 2008, 29(8): 979.
[11] Mohrman D E, Heller L J. Cardiovascular physiology [M]. New York: Mc-Graw-Hill Medical, 2002,
[12] Inan O T, Etemadi M, Wiard R M, et al. Robust ballistocardiogram acquisition for home monitoring[J]. Physiological Measurement, 2009, 30(2): 169.
Outlines

/