Control and Realization of Time Sequence for Low-dose Cone-Beam CT Imaging System

  • LUO Shouhua;CUI Yannan;BO Tengfei;LI Guang;CHEN Gong
Expand
  • 1. School of Biological Science and Medical Engineering, Southeast University, Nanjing 210096, China;2. Information Department, Jiangsu Province Hospital of Traditional Chinese Medicine, Nanjing 210029, China

Received date: 2012-01-31

  Revised date: 2012-02-05

  Online published: 2012-02-18

Abstract

The basic working principle of Cone Beam Computer Tomography (CBCT) is briefly described, the advantages and disadvantages of both continuous and pulsed X-ray imaging methods are analyzed, and the pulsed X-ray imaging method is adopted to implement the design of low-dose CBCT. On the basis of this, the internal and external trigger modes of PaxScan 2520 flat panel detector are realized. The experiment results show that the method of continuous X-ray imaging needs dozens of seconds for the process rotating a circle, in the meanwhile, the method of pulsed X-ray imaging only needs about four seconds for doing same job. In a word, pulsed X-ray imaging reduces the X-ray radiation time considerably and improves the safety performance of CBCT significantly. Besides, pulsed X-ray imaging with internal and external trigger modes has no difference in terms of image quality. The choice for the modes depends on whether or not needs active control for X-ray emission.

Cite this article

LUO Shouhua;CUI Yannan;BO Tengfei;LI Guang;CHEN Gong . Control and Realization of Time Sequence for Low-dose Cone-Beam CT Imaging System[J]. Science & Technology Review, 2012 , 30(5) : 23 -26 . DOI: 10.3981/j.issn.1000-7857.2012.05.002